Viasensor G150-11L Carbon Dioxide (0-10,000 ppm) Analyzer with O2 (0-100%) and RH Probe
G150-11L
QVT-2011010812355147595
Viasensor G150-11L Industrial Carbon Dioxide (0-10,000 ppm) Analyzer with O2 (0-100%) and RH probe monitors CO2 levels from 0-10,000 ppm and O2 from 0-100% for the site auditing, laboratory, and medical fields. Elevated CO2 levels can pose a health risk to employees and it is important to monitor.
The CO2 sensor is a dual wavelength IR Cell with a reference channel and features a quick response time. The O2 sensor is an electrochemical cell that has an approximate lifetime of 3 years in air. The Viasensor G150-11L has an internal pump with a typical flow rate of 100 cc/min. Ports for dual temperature probes is available with the probes sold separately. The LCD Display is easy to read and features a LED back light for working in darker environments.
The Viasensor G150-11L has a 1 year warranty. Made in the UK. Monitors CO2 from 0-10,000 ppm Monitors O2 from 0-100% Includes 15mm Diameter RH Probe Ports for dual temperature probes (probes sold separately) Comes with Owners Manual, Battery Charger, and Sample Tube Kit.
The Viasensor is ISO 9001 Certified Company.
The CO2 sensor is a dual wavelength IR Cell with a reference channel and features a quick response time. The O2 sensor is an electrochemical cell that has an approximate lifetime of 3 years in air. The Viasensor G150-11L has an internal pump with a typical flow rate of 100 cc/min. Ports for dual temperature probes is available with the probes sold separately. The LCD Display is easy to read and features a LED back light for working in darker environments.
The Viasensor G150-11L has a 1 year warranty. Made in the UK. Monitors CO2 from 0-10,000 ppm Monitors O2 from 0-100% Includes 15mm Diameter RH Probe Ports for dual temperature probes (probes sold separately) Comes with Owners Manual, Battery Charger, and Sample Tube Kit.
The Viasensor is ISO 9001 Certified Company.
Features
| Characteristics: | Viasensor G150-11L Industrial Carbon Dioxide (0-10,000 ppm) Analyzer with O2 (0-100%) and RH probe |
| IAQ analyser package 0-10,000 CO2 plus 0-100 percent O2 with internal pump and G1.8 RH probe | |
| Improved accuracy on CO2 readings | |
| Quick verification of CO2 levels for site audits | |
| Time saving with dual temperature probes | |
| Data logging for long-term application | |
| Easy to read large well-lit display | |
| Built-in gas moisture removal | |
| Includes: | Battery charger, sample tube/filters and operating manual |
Specifications
| Specifications | Viasensor G150-11L Carbon Dioxide (0-10,000 ppm) Analyzer with O2 (0-100%) and RH Probe | |
| Gas Ranges | ||
| Gases Measured | CO2 | By custom dual wavelength infra-red with reference channel |
| O2 | By Internal Electrochemical Cell | |
| Oxygen Cell Lifetime | Approximately 3 years in air | |
| Range | CO2 | 0 - 10,000 ppm |
| O2 | 0 - 100% | |
| Measurement Accuracy | Gas | |
| CO2 | ||
| Temperature dependence: ± 0.2% reading (typical at 5% CO2) | ||
| Pressure dependence: ± 0.02% of reading / hPa (typical at 5% CO2) | ||
| O2 | ± 1.0% Full Scale @ constant temperature and pressure | |
| ± 2.0% Full Scale over operating temperature range | ||
| Response time T90 | CO2 | = 20 seconds |
| O2 | = 60 seconds | |
| Environmental Conditions | ||
| Operating Temperature | 41 °F to 104 °F; 5 °C to 40 °C | |
| Relavtive Humidity | 0 - 95% non condensing | |
| Barometric Pressure | ± 201 "H20 from calibration pressure, (± 500mbar) | |
| Pump | ||
| Flow | 100 cc/min typically | |
| Power Supply | ||
| Battery Type | Li Ion | |
| Battery Life | 10 Hours (8 hours with pump) | |
| Battery Lifetime | > 300 Cylces | |
| Battery Charger | 5v DC external power supply and internal charging circuit | |
| Charge Time | 3 Hours | |
| Alternative Power | USB Connector | |
| DC Power Supply | ||
| Physical | ||
| Weight | 1.1 lbs (495 grams) | |
| Dimensions | 6 x 4 x 2 3/16 inches (165 x 100 x 55mm) | |
| Case Material | ABS / Polypropylene with Silicon Rubber Inserts | |
| Keys | 17 Resin capped Silicon Rubber Keys | |
| Display | Liquid Crystal Display, 128 x 64 pixel | |
| With RGB LED Backlight | ||
| Gas Sample Filters | Built-in gas dryer tube to remove moisture | |
| User replaceable PTFE water trap filter | ||
| Facilities | ||
| Temperature (Optional) | x 2 using optional probes, range 32 °F to 122 °F (0°C to +50.0 °C) | |
| Temperature accuracy, typical | ±0.2 °C from 32 to 44 °C, ±0.5 °C | |
| Barometric pressure | 800-1,200 mbar | |
| RH measurement (Optional) | RH probe 0-100% RH non condensing | |
| RH accuracy | ±1.5% RH across the range | |
| Visual and audible alarm | User selectable CO2 alarm levels | |
| Communications | USB type B mini-connect or, HID device class | |
| Data Storage | 1,000 reading sets + 270 events | |
Proposition 65 Warning for California Residents: This product can expose you to chemicals including Benzene, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov.