Kanomax 6501-CE Climomaster with Pressure Sensor and Analog Output
6501-CE
QVT-2012101771219974562
Kanomax 6501-CE Climomaster Anemometer with Analog Output and Pressure Sensor is a highly accurate hot-wire anemometer that simultaneously measures and displays air temperature, air velocity, flow rate, pressure, and humidity. The Kanomax anemometer features an ergonomic design with backlit display and anti-slip grip. Made in Japan. *Probes sold separately.
Features
| Characteristics: | Kanomax 6501-CE Climomaster Anemometer with Analog Output and Pressure Sensor |
| 2% air velocity accuracy: highest accuracy on the market | |
| Sleek, ergonomical design with an anti-slip grip and highly visible backlit display |
|
| Wide range of probe choices for challenging applications | |
| Research quality at a contractor-friendly price | |
| Equipped with an automatic atmospheric pressure compensation function | |
| Simultaneously measures and displays air velocity, air temperature, humidity, flow rate, and pressure | |
| Generous datalogging capacity (up to 20000 records) and the USB interface make it easy to upload to your PC or laptop | |
| Optional Windows 7 compatible software provides real-time graphical display of date for field analysis |
|
| Now tracks its own calibration cycle. It will now remind you when it is due for calibration |
|
| Optional hands-free carrying case is available | |
| Includes 79 in (200cm) probe cable, USB cable, 6 pcs. AA batteries, carrying case, and NIST Certificate |
|
| Warranty | 2 Years |
Specifications
| Specifications | Kanomax 6501-CE Climomaster Anemometer with Analog Output and Pressure Sensor | |
| Air Velocity | Range | 10 to 9840 FPM (0.01 to 50.0 m/s) *Varies by probe |
| Accuracy | ± 2% of reading or 0.015 m/s whichever is greater | |
| Resolution | 0.01 m/s (0.01 to 9.99 m/s), 0.1 m/s (10.0 to 50.0 m/s) | |
| Temperature | Range | -4 to 158°F (-20 to 70°C) |
| Accuracy | ± 1.0°F (0.5°C) | |
| Resolution | 1.0°F (0.1°C) | |
| Humidity | Range | 2.0 to 98.0 %RH *Varies by probe |
| Accuracy | ± 2.0 %RH | |
| Resolution | 0.1 %RH | |
| Pressure | Range | -5.00 to +5.00 kPa |
| Accuracy | ± (3% of reading + 0.01) kPa | |
| Resolution | 0.01 kPa | |
| Interface | Digital | USB and RS232C (for print-out) |
| Analog | DC 0 to 1 V (select one from Air Velocity, Temp, Humidity and Pressure) | |
| Data Storage Capacity | Up to 20,000 records | |
| Power Supply | 6 x AA batteries or AC Adapter | |
| Operation Range | Main Unit | 41 to 104°F (5 to 40°C) with no visible condensation |
| Probe (Optional) | -4 to 158°C (-20 to 70°C) with no visible condensation | |
| Storage | 14 to 122°C (-10 to 50°C) with no visible condensation | |
| Weight | 0.9lbs (400g) | |
| Dimensions | Main Body: 3.4 x 7.4 x 1.6 inches (88 x 188 x 41 mm) | |
| Warranty | 2 years | |
Proposition 65 Warning for California Residents: This product can expose you to chemicals including Benzene, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov.
Resources