Datasheet: Bacharach-ECA450 |
 |
| Bacharach 24-8400 ECA 450 Combustion Analyzer Kit Description: |
 Bacharach 24-8400 ECA 450 Combustion Analyzer Kit conducts accurate combustion and emissions tests that will help you meet environmental regulations and ensure your equipment is burning efficiently.

Bacharach 24-8400 ECA 450 measures Oxygen, Carbon Monoxide, Nitric Oxide, Nitrogen Dioxide and Sulfur Dioxide, as well as combustibles, temperature and draft. It also calculates combustion efficiency, excess air, carbon dioxide, NOx (Nitrogen Oxides) and pollution units.

Bacharach 24-8400 ECA 450 takes continuous samplings of furnaces, boilers and other industrial combustion equipment for up to eight hours and generates precise, reliable combustion and environmental measurements and calculations. And it generates this complex data easily — even at selected intervals — enabling plants to make more accurate equipment adjustments and better assess pollution levels.

Bacharach 24-8400 ECA 450 helps improve overall combustion efficiency, reduce fuel costs, and determine emissions compliance. Made in the USA. Free ground shipping within continental US.
 |
| Bacharach 24-8400 ECA 450 Combustion Analyzer Kit Characteristics: |
| Power: |
 Universal AC adapter and an internal battery pack.
 |
| Pumps and Probe: |
 Two pumps are included.
 |
| Size: |
 13.5"H x 18.5"W x 9"D (342.9 x 467.0 x 228.6 mm)
 |
| Weight: |
 25 lbs (11.34 kg)
 |
| Warranty: |
 1 year, extended warranty available
 |
| |
 ECA 450 NOx kit with conditioner.
 |
| Calculates: |
 Calculates and displays combustion efficiency, excess air, and CO2.
 |
| Measures: |
 Measures and displays O2 and CO in the flue gas, pressure (draft),
primary air temperature, and stack temperature.
 |
| |
 Stores over 1000 individual combustion and/or pressure test records,
which can later be recalled for viewing or printing.
 |
| |
 Stored test records can be downloaded to a personal computer and stored
as a text file, which can then be opened in a spreadsheet program for
analysis.
 |
| |
 Performs datalogging.
 |
| Optionally measures: |
 Optionally measures and displays NO, NO2, SO2, HC (methane equivalent
combustibles), and CO in the high range of 4,000 to 80,000 ppm).
 |
| Optionally calculates: |
 Optionally calculates NOx (the combination of NO and NO2), and
calculates CO, SO2, and NOx individually referenced to a user defined
Oxygen level of between 0 and 15%.
 |
| |
 Optionally displays pollution conversions for CO, NO, NO2, and SO2.
Pollution conversions include parts per million, pounds of pollutant
per million BTU, milligrams of pollutant per cubic meter of gas, and
grams of pollutant per gigajoule.
 |
| |
 Large, easy-to-read fluorescent display that is visible from wide
angles.
 |
| |
 EPA-ETV test verified performance.
 |
| |
 Modular design lets you add sensors as your needs change.
 |
| |
 Built-in printer and complete data downloading to a PC capability.
 |
| |
 Adapter will accept input voltages from 100 to 240V.
 |
| |
 The first pump supplies gas sample to the sensors.
 |
| |
 The second pump supplies fresh air to purge the low-range CO sensor when CO levels exceed 400 ppm.
 |
| |
 Fully charged battery pack provides a minimum of 8 hours of operation.
 |
| |
 Battery life of over eight hours.
 |
| Bacharach 24-8400 ECA 450 Combustion Analyzer Kit Specifications: |
| Measured Data: |
 Oxygen: 0.1 to 20.9%
 |
| |
 Carbon Monoxide (hydrogen compensated): 0 to 4000 ppm
 |
| |
 Carbon Monoxide High: 4001 to 80,000 ppm
 |
| |
 Nitric Oxide*: 0 to 3500 ppm
 |
| |
 Sulfur Dioxide*: 0 to 4000 ppm
 |
| |
 Combustibles*: 0 to 5% (application dependent)
 |
| |
 Stack Temperature: +- 4 to 2400 F (+- 20 to 1315 C)
 |
| |
 Primary/Ambient Temperature: +- 4 to 999 F (+- 20 to 999 C)
 |
| |
 Pressure/Draft: +- 27.7 to 27.7 inches of Water
 |
| Calculated Data: |
 Combustion Efficiency: 0.1 to 100.0%
 |
| |
 Excess Air: 1.0 to 250%
 |
| |
 Carbon Dioxide(dry basis): 0 to fuel dependent maximum
 |
| |
 NOx: 0 to 4000 ppm
 |
| |
 NOx (ref. to %O2): 0 to 17,000 ppm
 |
| |
 CO (ref. to %O2): 0 to 99,9999 ppm
 |
| |
 NO2 (ref. to %O2): 0 to 2100 ppm
 |
| |
 SO2 (ref. to %O2): 0 to 17,000 ppm
 |
| |
 Probe includes a standard probe and hose assembly equipped with a water trap, particulate filter, probe stop, 15 feet of hose, and 12-inch probe tube.
 |