Datasheet: PM53A |
 |
| Amprobe PM53A Digital Multimeter Description: |
 Amprobe PM53A is fully autoranging and has VolTect, a built-in non-contact AC voltage detector. Plus it uses AutoTect to automatically detect the measurement of AC volts, DC volts or resistance. Safety rated to Cat III 300V, CAT II 450V, cULus listed.
 |
| Amprobe PM53A Digital Multimeter Characteristics: |
| |
 Unique AutoTect feature directs the meter to know if it is measuring AC volts, DC volts or resistance with no user interface required
 |
| |
 The AutoTect feature has lower input impedance, which zeros out ghost voltages
 |
| |
 Measures capacitance, frequency, continuity with beeper
 |
| |
 Includes VolTect, a built-in, noncontact AC voltage detector
 |
| |
 Complete user protection with full safety ratings of CAT III 300 V, CAT II 450 V and cULus listing
 |
| |
 Large oversize digital display
 |
| |
 Protective silicon holster included, attached test leads, battery and manual
 |
| |
 One-year No Hassle warranty
 |
| Amprobe PM53A Digital Multimeter Specifications: |
| DC Voltage |
 Range 400.0 mV
Accuracy +/-(1.0 %+2 dgt)
Range 4.000 V, 40.00 V, 400.0 V
Accuracy +/-(2.0 %+2 dgt)
Range 600 V
Accuracy +/-(2.5 %+4 dgt)
NMRR > 50 dB @ 50 Hz/60 Hz
CMRR > 120 dB @ dc, 50 Hz/60 Hz; Rs=1 kΩ
Input impedance 10 MΩ, 30 pF nominal; (1000 MΩ for 400.0 mV range)
 |
| AC Voltage |
 Range 50 Hz - 60 Hz 4.000 V, 40.00 V, 400.0 V
Accuracy +/-(2.0 %+5 dgt)
Range 60 Hz - 500 Hz 4.000 V, 40.00 V, 400.0 V
Accuracy +/-(3.0 %+5 dgt)
Range 50 Hz - 500 Hz 600 V
Accuracy +/-(3.5 %+5 dgt)
Input impedance 10 MΩ, 30 pF nominal
CMRR > 60 dB @ dc to 60 Hz, Rs=1 kΩ
 |
| Capacitance |
 Range 500.0 nF, 5.000 μF, 50.00 μF, 500.0 μF, 3000 μF
Accuracy +/-(3.5 %+6 dgt)
 |
| Resistance |
 Range 400.0 Ω
Accuracy +/-(1.5 %+6 dgt)
Range 4.000 kΩ, 40.00 kΩ, 400.0 kΩ
Accuracy +/-(1.0 %+4 dgt)
Range 4.000 MΩ
Accuracy +/-(1.5 %+4 dgt)
Range 40.00 MΩ
Accuracy +/-(2.5 %+4 dgt)
Open circuit voltage 0.4 V DC typical
 |
| Frequency |
 Range 400 Hz, 4 kHz, 40 kHz, 400 kHz, 1 MHz
Accuracy +/-(0.5 %+4 dgt)
Sensitivity (V-peak)
5 Hz - 100 kHz > 1.3 Vp 100 kHz - 500 kHz > 2.2 Vp 500 kHz - 1 MHz > 4.2 Vp
 |
| Diode Tester |
 Test current (typical) 0.25 mA
Open circuit voltage 1.6 V dc
 |
| Audible Continuity Tester |
 Open circuit voltage 0.4 V dc typical
Audible threshold 65 Ω +/-55 Ω
 |

Need Technical Advise? Please Call: 602-795-4033