| Datasheet: |
 |
| Amprobe Meterman AC60 Clamp Meter Description: |
 Amprobe Meterman AC60 Clamp Meter is a durable AC clamp multimeter with a safe hand guard design. A good feature is it's Power Off 30 minutes after no function change. Discontinued product.
 |
| Amprobe Meterman AC60 Clamp Meter Characteristics: |
| |
 Professional's 600 A, AC clamp meter
 |
| |
 Autoranging for fast response
 |
| |
 AC voltage to 600V
 |
| |
 Data Hold button
 |
| |
 Continuity Beeper
 |
| |
 Jaw Size 40mm (1.6")
 |
| |
 4000-count
 |
| |
 Carrying case included
 |
| |
 CAT II 600 V, CAT III 300 V rated
 |
| Amprobe Meterman AC60 Clamp Meter Specifications: |
| AC Voltage ( 40 @ 500 Hz) |
 Ranges 400 V, 600 V Ranges 400 V, 600 V
Accuracy +/-(1.2 % + 5 dgts)
Resolution l00 mV in 400 V range
Input Impedance 10 MΩ // < 100 pF
 |
| AC Current (50/60 Hz) |
 Range 400 A, 600 A
Accuracy 400 A +/-(1.9 % rdg + 5 dgts)
Accuracy 600 A +/-(2.9 % rdg + 5 dgts)
Resolution 0.1 A in 400 A range
AC Conversion Type Avg. sensing indicating
 |
| Resistance |
 Range 4 kΩ, 40 kΩ
Accuracy +/-(2.0 % rdg + 9 dgts)
Resolution 1 Ω in 4 kΩ range
Max Open Circuit Voltage 1 V
 |
| Continuity |
 Buzzer sounds when resistance is < 200 Ω
 |
| Overload Protection |
 AC Volts, Resistance 600 V rms
AC Current 800 A rms
 |
| Power |
 Single 9V battery (NEDA 1604,IEC 6F22)
 |
| Display |
 3 3⁄4 Digit LCD, 3999 counts
 |
| Display Update Rate |
 4/sec
 |
| Battery Life |
 Alkaline 200 hours
 |
| Dimensions (WxLxH) |
 88 x 215 x 36 mm (3.5˝ x 8.5˝ x 1.4˝)
 |
| Weight |
 320 g (11.3 oz)
 |
| Operating Temperature |
 0 to +45 Degrees C, < 75 % R.H.
 |
| Storage Temperature |
 -20 to +60 Degrees C < 80 % R.H.
 |
| Temperature Coefficient |
 0.2 x (spec. Acc'y)/ Degrees C, < 18 Degrees C or >28 Degrees C
 |

Need Technical Advise? Please Call: 602-795-4033
Browse Similar Products |